Cardiac device infection risk (BLISTER)

Estimates the 12-month infection risk after implanting or revising a pacemaker, ICD or resynchronisation device.

Updated 31 October 2026

When to use it

Before or at the end of a transvenous cardiac implantable device procedure, to decide on preventive measures such as an antibacterial envelope.

What it needs

  • eGFR < 30 mL/min/1.73 m² at the time of the procedureYes or no
  • CRP ≥ 50 mg/L within the previous 24 hYes or no
  • Pocket open for ≥ 120 minutesYes or no
  • Type of procedureOptions: Pacemaker implant or generator change · ICD implant or generator change · CRT implant or generator change · New lead inserted or lead revised without extraction · Lead extraction
  • Immunosuppression (steroids, immunosuppressants or disease such as HIV)Yes or no
  • AgeOptions: < 60 years · 60–69 years · ≥ 70 years
  • Intervention on the same pocket within the previous 2 yearsYes or no
  • Previous device proceduresOptions: None (first procedure) · 1 previous · ≥ 2 previous

The points for each answer and the automatic total are in the app.

How to interpret it

BLISTER score
ResultInterpretation
≤ 5< 6: low risk; antibacterial envelope not cost-effective
≥ 6≥ 6: high risk; consider an antibacterial envelope

Practical tip

In the derivation study, a BLISTER score ≥ 6 (≈ 11 % of procedures) was the cost-effective threshold for an antibacterial envelope.

2 more practical tips in the app.

Reference

A support tool for healthcare professionals: it guides and does not replace clinical judgement.