Cardiac device infection risk (BLISTER)
Estimates the 12-month infection risk after implanting or revising a pacemaker, ICD or resynchronisation device.
Updated 31 October 2026
When to use it
Before or at the end of a transvenous cardiac implantable device procedure, to decide on preventive measures such as an antibacterial envelope.
What it needs
- eGFR < 30 mL/min/1.73 m² at the time of the procedure
- CRP ≥ 50 mg/L within the previous 24 h
- Pocket open for ≥ 120 minutes
- Type of procedure
- Immunosuppression (steroids, immunosuppressants or disease such as HIV)
- Age
- Intervention on the same pocket within the previous 2 years
- Previous device procedures
The points for each answer and the automatic total are in the app.
How to interpret it
| Result | Interpretation |
|---|---|
| ≤ 5 | < 6: low risk; antibacterial envelope not cost-effective |
| ≥ 6 | ≥ 6: high risk; consider an antibacterial envelope |
Practical tip
In the derivation study, a BLISTER score ≥ 6 (≈ 11 % of procedures) was the cost-effective threshold for an antibacterial envelope.
2 more practical tips in the app.
Reference
A support tool for healthcare professionals: it guides and does not replace clinical judgement.