Qp/Qs ratio by Doppler echocardiography
Ratio of pulmonary to systemic flow to quantify an intracardiac shunt.
Updated 30 October 2026
When to use it
Atrial or ventricular septal defect or patent ductus, to assess the haemodynamic impact of the shunt.
What it needs
- RVOT diameter
- RVOT velocity–time integral (pulsed-wave)
- LVOT diameter
- LVOT velocity–time integral (pulsed-wave)
Formula
Qp = π × (RVOT d / 2)² × RVOT VTI
Qs = π × (LVOT d / 2)² × LVOT VTI
Qp/Qs = Qp / Qs
How to interpret it
| Result | Interpretation |
|---|---|
| < 1.5 | Small or no shunt (< 1.5) |
| 1.5 to < 2 | Significant shunt (1.5–2) |
| ≥ 2 | Large shunt (≥ 2) |
Practical tip
A Qp/Qs ≥ 1.5 is usually considered significant and, with right ventricular dilatation, supports closing the defect.
2 more practical tips in the app.
Reference
- Kitabatake A, Inoue M, Asao M, et al. Noninvasive evaluation of the ratio of pulmonary to systemic flow in atrial septal defect by duplex Doppler echocardiography. Circulation. 1984;69(1):73-79.
- Silvestry FE, Cohen MS, Armsby LB, et al. Guidelines for the echocardiographic assessment of atrial septal defect and patent foramen ovale: from the American Society of Echocardiography and Society for Cardiac Angiography and Interventions. J Am Soc Echocardiogr. 2015;28(8):910-958.
A support tool for healthcare professionals: it guides and does not replace clinical judgement.